Calibrated, certified, and patient ready with a 90 day warranty
The Abbott Plum A+ Infusion Pump is a general-purpose infusion device that provides precise delivery of multiple therapies across the general spectrum of clinical care.
The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With its primary line, secondary line, and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories and the LifeShield needleless protection systems makes the Plum A+ convenient and cost-effective infusion system.
Therapy modes include:
- Multistep Programming
- Loading Dose
- Dose Calculation
The Plum A+are designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types.
Both are designed to be used in most areas of patient care, including, but not limited to:
- General Floor
- Labor/Delivery/Post Partum
- Burn Unit
- Medical/Surgical
- OR/Anesthesia
- Hemodialysis
- ICU/CCU
- Post Op/Recovery
- Oncology
- Pediatrics
- Cardiac Cath Lab
- Mobile Intensive Care
- Neonatolgy
- Emergency
- Nutritional
Specifications
- Dimensions: 8" H x 8" W x 6" D excluding pole clamp
- Weight: 9.5 lbs. with battery
- Power: 120V, 50-60 Hz, 35VA
- Battery: One sealed lead acid, rechargeable 6V battery, internal
Operating Range
Delivery Rate Range
Lines A & B:
- 0.1-99.9 ml/hr in 0.1 ml increments
- 100-999 ml/hr in 1 ml increments
Concurrent Delivery: 0.5 ml/hr minimum for each line
Plum Set: 500 ml/hr cumulative (A+B) maximum
KVO Rate: 1.0 ml/hr or the last primary delivery rate, whichever is less
VTBI Range: 0.1-99.9 ml in 0.1 ml increments 100-9999 ml in 1 ml increments
Line Programming Options
- Therapy Selection
- Nurse Call Back
- Delayed Start Setting
- Drug Label Library
- Concurrent delivery
- Titration
- Automated Secondary drug delivery (Piggyback)
Plumset Capabilities
- Anti Free-Flow Protection
- Direct Connection for syringe delivery
Air Management
- Air Trap
- Air Removal/Backpriming
- Air Detection-Proximal
- Air Detection-Distal
Other Features
- Nonpulsatile volumetric accuracy
- Microprocessor control
- Large liquid crystal display (LCD) screen
- Panel back illumination on mains power
- Lockout switch
- Standard fullfill, partfill, syringe and vial use
- Parenteral and nonparenteral (enteral) fluid delivery
- Blood and blood products delivery